Chromatography Liquid Chromatography (LC) Gas Chromatography (GC) Thin-layer Chromatography(TLC)
Spectrometer AAS ICP Raman Spectrometer FT-IR Spectrometer UV-Vis Spectrophotometer Terahertz Spectrometer Infrared Spectrometer Terahertz Imager Spectroscopy Consumables Others
Mass Spectrometer MALDI-TOF-MS GC-MS ICP-MS
Physical Property Analysis Instrument Differential Scanning Calorimeter Melting Point Apparatus Refractometer Polarimeter Viscometer Density Meter
Sample Preparation Laboratory Water Purification System Evaporator Extractor Centrifuge Mixer Pipette Digester Others
Life Science Safety Cabinet Freezer PCR Microplate Reader Incubator Others
Elemental Analysis Kjeldahl Apparatus TOC Analyzer Carbon&Sulphur Analyzer Metal Analyzer Others
Lab Consumables Biochemical Consumables SPE Cartridges QuEChERS Kits
PPE Instruments Oxygen Concentrator Sterilizer
Environmental Measurement Iinstruments Gas Measurement Device
Others Magnetic Content Analyzer Glassware Washer Capping Machine Freeze Dryer Burette PH Meter Calorimeter Titrator Nitrogen Generator Automatic Drop Point and Softening Point Tester
Digestion furnace is a type of sample digestion and conversion equipment developed according to the classical wet digestion principle. It is mainly used in agriculture, forestry, environmentalprotection, geology, petroleum, chemical and food industries as well as universities and scientific research institutes for digestion treatment of plant, seed, feed, soil and ore before chemical analysis.It is the best supporting product for Kjeldahl nitrogen analyzer.
Product Features
Infrared radiation technology is used for this series of products and stainless steel heating pipe is used for the heating element, which has the characteristics of fast heating and long life.
erospace coating technology is adopted for the surface of L series aluminum heating module, beautiful, durable and the design temperature is 450℃.
LCD display for the temperature controller with timed shutdown and alarm functions. Intelligent PID control technology is used with high temperature control accuracy and small temperatureimpact range, simple and easy to learn.
he instrument table top and test tube rack are welded with brushed stainless steel plates to prevent corrosion.
The optional exhaust system is corrosion and temperature-resistant, so that SO2 and other harmful gases escaping from the digestion tube can be discharged from the water into the sewer through the sewage pipe and the suction pump to effectively inhibiting the escape of harmfu gases.
Number of sample holes: 20
Pore size: Φ43.5mm
Heating module: Aluminium alloy
Design temperature: 400℃
Temperature control accuracy: ±1℃
Heating rate: ≈ 8-15°C/min
Timed shutdown: 1-999min
Operating voltage: AC220V/50Hz
Heating power: 2.8KW